[Multiplex Assay Kit for Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay)]

 

 

  • Product No.LML924Hu
  • Organism SpeciesHomo sapiens (Human) Same name, Different species.
  • Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
  • Test MethodDouble-antibody Sandwich
  • Assay Length3.5h
  • Detection Range0.05-50ng/mL
  • SensitivityThe minimum detectable dose of this kit is typically less than 0.017 ng/mL.

Specificity

This assay has high sensitivity and excellent specificity for detection of Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)84-10495
EDTA plasma(n=5)96-10599
heparin plasma(n=5)79-8885

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Semaphorin 5A (SEMA5A) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)87-95%88-102%95-102%95-104%
EDTA plasma(n=5)92-101%96-104%97-105%94-102%
heparin plasma(n=5)98-105%93-101%82-96%86-102%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

 


 